A guide to Differential Interference Contrast (DIC)

A guide to Differential Interference Contrast (DIC)

This guide provides an overview of Differential Interference Contrast (DIC), how DIC works and how to set DIC up on an upright microscope - Scientifica SliceScope.

Differential Interference Contrast (DIC) is a microscopy technique that introduces contrast to images of specimens which have little or no contrast when viewed using brightfield microscopy. The images produced using DIC have a pseudo 3D-effect, making the technique ideal for electrophysiology experiments.

Using DIC produces high resolution images with good contrast. It is best for visualising unstained samples. When viewing fluorescently labelled compounds, using DIC reduces the fluorescence intensity slightly and so decreases the quality of fluorescence images. However, this decrease is not significant.

As an Infrared (IR) light source is usually used, DIC is ideal for imaging thick specimens such as brain slices, as IR light penetrates deeper into tissues than visible light, due to its longer wavelength.

An advantage of using DIC over other contrast techniques, such as phase contrast or oblique contrast is that in DIC the full aperture of the microscope is used. For example, in phase contrast, the annulus of the condenser restricts the aperture, reducing the resolution of the image. Unlike phase contrast, DIC images are not disturbed by halo artefacts.

How does Differential Interference Contrast work?

In DIC, light emitted from the source is linearly polarised by passing through a polariser. The linearly polarised beam of light enters an objective-specific prism, which splits it into two rays that vibrate perpendicular to each other. The rays are parallel as they pass through a condenser, but as they are vibrating perpendicular to each other, they are unable to cause interference.

The split beams pass through the specimen. The specimen’s varying thickness and refractive indices alter the wave paths of the beams. They then enter the objective, where they are focussed above the rear focal plane. The two beams enter a second prism, in the nosepiece, which combines them. Because the beams passed through different parts of the specimen, they have different lengths.

The analyser, which is a second polariser, brings the vibrations of the beams into the same plane and axis, causing destructive and constructive interference to occur between the two wavefronts. The light then travels to the eyepiece or camera, where a DIC image with differences in intensity and colour, can be seen.

A diagram showing the journey of light in DIC
A diagram showing the journey of light in DIC. The yellow lines represent non-polarised light, whilst red lines represent polarised light.


How to set up Differential Interference Contrast

DIC can be utilised on virtually any upright or inverted microscope, including the Scientifica SliceScope, as long as the polariser, condenser and prisms can be installed.


The Scientifica SliceScope with DIC components labelled
The Scientifica SliceScope with DIC components labelled

Before setting up DIC on a microscope, ensure all components are free of dust and debris, as dirty objective, condenser and lens elements can reduce image quality.

1. If using a white light source, insert an infrared bandpass filter into the slider on the SliceScope filter wheel and secure in place with a locking screw, this ensures only infrared light passes through to the polariser. If using an IR-LED, a bandpass filter is not needed.

2. Insert the polariser into the silver-plated section of the filter wheel, underneath the condenser, shown in the image below. The polariser is on a rotatable mount, enabling it to be rotated when aligning DIC.


The polariser inserted in the filter wheel
The polariser inserted in the filter wheel


3. Ensure your condenser does not have a protective covering on it, as this will obscure images produced using DIC.

4. Keeping the condenser facing upwards, use a hex key to unscrew the two screws either side of the condenser to open it up. Turn the top section only upside down.

5. Loosen the grub screw on the side of the condenser and remove the spacer ring from the top half, as this is not needed. Insert the objective-specific prism, lining up the metal pin on the prism with the slot in the condenser. Tighten the grub screw and put the condenser back together. Each objective requires a different objective-specific prism, based on the magnification of the objective, for example a 40x objective requires a WI-DIC40HR objective specific prism. Usually, a prism is only required for the high magnification objective you are using.


The WI-DIC60HR prism in the condenser and the condenser setup on the SliceScope with the filter wheel below.
The left image shows the WI-DIC60HR prism in the condenser. On the right is the condenser setup on the SliceScope with the filter wheel below.

6. The second DIC prism (WI-DICT or WI-DICTHRA) is in a fixed mount which is inserted into the nosepiece above the objective. Loosen the grub screw on the right-hand side of the nosepiece arm and slide the nosepiece (WI-SRE) forward. Insert the second DIC prism into the nosepiece and secure it using the small screw at the rear-right corner of the nosepiece.

If using a MOC, pictured below, loosen the four screws to open it and then follow the instructions above.


The objective prism in the MOC and the nosepiece, where the objective prism is located.
The image on the left shows the objective prism in the MOC. The image on the right is the nosepiece, where the objective prism is located.

7. If fluorescence is being used, insert the analyser (U-AN) into the fluorescence turret, as shown below. If fluorescence isn’t being used, insert the analyser (U-ANT) into the U-KPA.

8. The U-KPA can then be secured on the SliceScope under the imaging lens or trinocular head.

9. The polariser and analyser need to be aligned so that their transmission pathways cross at a 90 degree angle to one another. This can be achieved by rotating the polariser in the filter wheel. As this can be rotated 360 degrees, the analyser and U-KPA can be mounted in any orientation.


U-AN analyser, and the U-AN analyser inserted into a fluorescent turret on the SliceScope
The image on the left is a U-AN analyser, and the image on the right shows the U-AN analyser inserted into a fluorescent turret on the SliceScope.

10. For DIC microscopy to be performed, Koehler illumination must be first set-up. Read our 6 step guide to Koehler illumination to help you set this up.

11. Once Koehler illumination has been set-up, DIC will need to be aligned.

Aligning Differential Interference Contrast

1. Focus on a blank sample plate using either a 4x or 10x objective.

2. Move the DIC slider with the U-AN analyser into the light path. The U-ANT is not on a slider and is in a fixed position which is always in the light path.

3. If using a trinocular head, remove one eyepiece and view the sample directly down the trinocular head.

4. a. When using eyepieces: rotate the polariser until there is a dark stripe through the centre of the field of view. This indicates that the polariser and analyser are aligned at 90 degrees to each other.

b. When using a camera: rotate polariser until the image is at its darkest.

5. If the condenser and objectives were removed, put them back at this point. Double-check that your condenser is still set-up for Koehler illumination.

6. Use your DIC objective to bring your sample into focus.

7. Add a specimen so that contrast can be optimised for the sample you are viewing.

8. As the SliceScope uses de Sénarmont DIC, a quarter waveplate is required, shown in the image below. The quarter waveplate introduces a bias retardation. Adjusting this changes whether structures appear concave or convex. Unscrew the locking knob and adjust the quarter wave plate as desired.

9. Lock the quarter waveplate in position using the locking knob.


Zoomed in section of the SliceScope with DIC components labelled
Zoomed in section of the SliceScope with DIC components labelled

10. The lever in the image above controls the condenser iris aperture. If using eyepieces, remove one eyepiece and look down so you can see the internal condenser iris. Adjust the lever until the condenser iris just disappears from the field of view. When using a camera, adjust the lever until the image is at its sharpest.

11. DIC is a function of specimen orientation. Therefore, rotating the sample can improve or change the contrast of the image.

You should now have a clear DIC image! An example of a DIC image is below; the image has good contrast and the C.elegan has a 3D effect.

An adult Caenorhabditis elegans nematode imaged using DIC by Bob Goldstein
An adult Caenorhabditis elegans nematode imaged using DIC by Bob Goldstein

If you have any questions or require further guidance, please feel free to contact us at [email protected].

Scientifica SliceScope

A stable and compact upright microscope designed to meet the demands of complex electrophysiology studies and related techniques - including two-photon imaging and optogenetics.

Scientifica have created a printable PDF guide for microscope users. Download it here:

Differential Interference Contrast Guide

Learn about other contrast techniques

Find out about Scientifica's latest product releases, company news, and developments through a range of news articles, customer interviews and product demonstration videos.

Sign up to receive our latest news

Contact Form

Contact us

* denotes required field

Select your interests