items in enquiry
Scientifica Multiphoton Detection Unit XL (MDU XL)
Scientifica Multiphoton Detection Unit XL (MDU XL)
Schematics
MDU XL Top View
MDU XL Side View
MDU XL schematics
Design and specifications
- Collection Optics: Primary Detection Dichroic: 60 x 40 x 1
- : Collection lens diameter: 45 mm
- : Maximum collection angle: ~±8° assuming 20mm objective back aperture
- : IR Blocking Filter: 680nm Shortpass 50mm Ø
- Emission Filters: Emission filter: 32mm Ø - one for each detector
- : Short wavelength perpendicular to the incident light, longer wavelength on the incident direction
- : Blocking filter: 32mm Ø for stimulation wavelength
- Photomultiplier Tubes: Two locally mounted detectors of the same type: GaAsP protected photomultiplier tube (H10770PA-40) or GaAsP photomultiplier tube with gating function (H11706P-40)
- : Signal output Connector: SMA
- : Gate input connector (gated GaAsP only): BNC
- Protection Circuitry: Protected GaASP: Integrated protection circuitry to trip PMT when overload detected
- : Gated GaAsP: Voltage gating functionality to avoid overload situations
- : All systems: Ambient light detector
- Objective and Nosepiece Compatibility: Objectives: M32, M27, M25, RMS threaded objectives (<20mm back aperture)
- : Nosepieces: Ships with single objective changer with 2 locking sliders
- : Compatible with: WI-SRE3 adapter (RMS only) or standard/DIC objective sliders
- Preamplifier Bandwidth: Options include: 20 MHz (resonant system), 1 MHz (galvo), or variable bandwidth all with low noise linear power supply
- Output Connection: BNC
- PMT Voltage Control: Manual or Software Control
- PMT Protection: Integrated protection circuitry in H10771P-40 only. For all PMT options, a remote ambient light sensor is used